Programming
463
Data Type Boolean type (Boolean)
Range ON|OFF|1|0
Preset Value ON
Query Response
{1|0}<newline><^END>
Examples
Dim Var as Interger
Ana.WriteString ":DISP:WIND1:TRAC1:STAT ON", True
Ana.WriteString ":DISP:WIND1:TRAC1:STAT?", True
Var=Ana.ReadNumber
Related Commands
:DISP:WIND:TRAC:MEM
Equivalent Softkey
Display > Display > Data | Data & Mem | OFF
COM Command Syntax (Internal Control Only)
SCPI.DISPLAY.WINDOW(Ch).TRACE(Tr).STATE
:DISP:WIND<Ch>:TRAC<Tr>:X:RLEV
Type
Command/Query
Syntax
:DISPlay:WINDow<Ch>:TRACe<Tr>:X[:SCALe]:RLEVel <Value>
:DISPlay:WINDow<Ch>:TRACe<Tr>:X[:SCALe]:RLEVel?
Description
This command sets/gets the value of the reference division line for X-axis.
Variable
Parameter Value
Description Value of reference division line
Data Type Numeric type (Real)