E4990A Help
472
:DISPlay:WINDow<Ch>:TRACe<Tr>:Y:TRACk:MODE?
Description
This command selects reference offset tracking method.
Variable
Parameter Selection Option
Description Reference offset tracking
Data Type Character string type (String)
Range OFF|PEAK|FREQuency
Preset Value OFF
Query Response
{OFF|PEAK|FREQ}<newline><^END>
Examples
Dim Var as String
Var= "peak"
Ana.WriteString ":DISP:WIND1:TRAC1:Y:TRAC:MODE" & Var, True
Ana.WriteString ":DISP:WIND1:TRAC1:Y:TRAC:MODE?", True
Var=Ana.ReadString
Related Commands
:DISP:WIND:TRAC:Y:TRAC:FREQ
Equivalent Softkey
Scale > Reference Tracking > Tracking > OFF|Track Peak|Track Freq
COM Command Syntax (Internal Control Only)
SCPI.DISPLAY.WINDOW(Ch).TRACE(Tr).Y.TRACK.MODE
:DISP:WIND<Ch>:X:SPAC
Type
Command/Query
Syntax
:DISPlay:WINDow<Ch>:X:SPACing {LINear|OBASe}
:DISPlay:WINDow<Ch>:X:SPACing?