Programming
513
COM Command Syntax (Internal Control Only)
SCPI.SENSE.ADAPTER.TYPE
:SENS<Ch>:ADAP:CORR:COLL
Type
Command
Syntax
:SENSe:ADAPter:CORRection:COLLect[:ACQuire]
{OPEN|SHORt|LOAD|PHASe}
Description
This command measures the adapter setup standard.
Variable
Parameter Selection Option
Description
Adapter setup standard
Data Type Character string type (String)
Range
OPEN: Open
SHORt: Short
LOAD: Load or E4990-61001
PHASe: Phase
Examples
Dim Var as String
Var= "OPEN"
Ana.WriteString ":SENS:ADAP:CORR:COLL " & Var, True
Related Commands
:SENS:ADAP:CORR:COLL:SAVE
Equivalent Softkey
Cal > Adapter Setup > Phase
Cal > Adapter Setup > Open
Cal > Adapter Setup > Short
Cal > Adapter Setup > Load
Cal > Adapter Setup > E4990-61001
Instrument Setup > Adapter Setup > Phase
Instrument Setup > Adapter Setup > Open
Instrument Setup > Adapter Setup > Short
Instrument Setup > Adapter Setup > Load