524
Equivalent Softkey
Cal > User Cal > Define Value > Open Cap (C)
Cal > Fixture Compen > Define Value > Open Cap (C)
COM Command Syntax (Internal Control Only)
SCPI.SENSE(Ch).CORRECTION(Type).CKIT.OPEN.C
:SENS<Ch>:CORR<Type>:CKIT:OPEN:G
Type
Command/Query
Syntax
:SENSe<Ch>:CORRection<Type>:CKIT:OPEN:G <Value>
:SENSe<Ch>:CORRection<Type>:CKIT:OPEN:G?
Description
This command sets/gets the open conductance value for the user
calibration and fixture compensation.
(<type>: User calibration=1, Fixture Compensation=2)
Variable
Parameter Value
Description Open conductance value
Data Type Numeric type (Real)
Range -1M ~ 1M
Preset Value
0
Unit S
Resolution 1p
Query Response
{numeric}<newline><^END>
Examples
Dim Var as Double
Var= -1000000
Ana.WriteString ":SENS1:CORR1:CKIT:OPEN:G " & Str(Var), True
Ana.WriteString ":SENS1:CORR1:CKIT:OPEN:G?", True
Var=Ana.ReadNumber