E4990A Help
534
Cal > User Cal > Execute Cal > Short
Cal > Fixture Compen > Short (x Ω x H)
COM Command Syntax (Internal Control Only)
SCPI.SENSE(Ch).CORRECTION(Type).COLLECT.ACQUIRE.SHORT
:SENS<Ch>:CORR<Type>:COLL:CLE
Type
Command
Syntax
:SENSe<Ch>:CORRection<Type>:COLLect:CLEar
Description
This command clears the measured data.
Examples
Ana.WriteString ":SENS1:CORR1:COLL:CLE", True
Related Commands
:SENS:CORR:COLL:ACQ:LOAD
:SENS:CORR:COLL:ACQ:OPEN
:SENS:CORR:COLL:ACQ:SHOR
Equivalent Softkey
Cal > User Cal > Execute Cal > Cancel > OK
COM Command Syntax (Internal Control Only)
SCPI.SENSE(Ch).CORRECTION(Type).COLLECT.CLEAR
:SENS<Ch>:CORR:COLL:FPO
Type
Command/Query
Syntax
:SENSe<Ch>:CORRection:COLLect:FPOints {FIXed|USER}
:SENSe<Ch>:CORRection:COLLect:FPOints?
Description
This command selects the frequency point type for both user calibration
and fixture compensation. The same type is applied to both user calibration
and fixture compensation.
Variable
Parameter Selection Option
Description Frequency point type
Data Type Character string type (String)