Measurement
85
The number of sweep points can be set to any
integer from 2 to 1601. When list is selected as a sweep type
(list sweep), use a list sweep table to set the number of points.
Setting the number of points is commonly applied to traces 1
and 2.
Selecting Sweep Direction
1. Press Sweep Setup to display the Sweep Setup menu.
2. Press Directions and select either Up or Down sweep direction.
The sweep direction enables you to select the
direction of the sweep. By default, the direction is Up. Down
direction refers to sweep of opposite direction (from right to
left).
Setting Time Delay for Measurement
Time delay can be set for the period before sweep or actual measurement
starts after the measurement signal is applied to the DUT. This function is
useful, for example, when a certain period is required before the
characteristics of the DUT become stable after the signal is applied.
Another application of this function is to observe changes in the impedance
of the DUT in the time domain for a long span.
Setting with Time Delay at Measurement Point
The time delay for each measurement points can be set by either or both
of following two methods. The time delay is defined as follows. The sweep
time in softkey is not the total sweep time.
Point Wait Time = Point Delay + (Sweep Time/Number of points)
Setting with Point Delay
1. Press Sweep Setup > Point Delay and input the desired time delay at
a point.
Setting with Sweep Time
1. Press Sweep Setup > Sweep Time and input the desired time.
Timing Chart for Sweep