108
KMHA 500-000099
Revision D
KDR
TM
AU-DDR Advanced U-Arm System with
Dynamic Digital Radiography (KDR AU DDR)
5. Click calculate.
6. The EI Calibration form will update.
a. “Expected EI Value” of the calibrated dosimeter in is units of EI (=100x μGy).
b. “Calculated EI Value” of the xed center ROI for the current connected panel is in units of EI when Scaler is
OFF.
c. “Scaler Value” is the “Expected EI Value” of the calibrated dosimeter divided by the “Calculated EI Value” of
the xed center ROI.
d. “Reported EI Value” is the “Scaler Value” multiplied by the “Calculated EI Value”.
7. Click ‘Save’ button to record all the parameters of EI calibration for the calibrated panel.