II. SAFETY
6.3 Emergency Stop
The emergency stop is located on the outer right corner of the bucky assembly. If the stand is not working properly or
making unintended movements, the emergency stop switch can be pressed to immediately stop the stand from moving.
The emergency stop only disables motorized mechanical movements of the stand and does not affect the generator and
other components of the system. To recover movement capability, make sure the area around the stand is clear and free of
obstructions, then pull the switch directly out and rotate it clockwise. Contact your service engineer if there are persistent
erratic movements in the stand.
Figure 6.3.1
Emergency Stop
6.4 Radiation and X-Ray Tube
WARNING! Make sure that patients, operators, and third parties are protected against unnecessary X-ray radi-
ation according to local regulations.
WARNING! The collimator lter must be veried so that the correct lter is used during exposure.
NOTE: Verbal and visual communication must be possible between the operator and the patient when expo-
sures are performed.
NOTE: The X-ray beam must never be outside the boundaries of the image receptor holder.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Konica Minolta KDR AU DDR and is the answer not in the manual?

Konica Minolta KDR AU DDR Specifications

General IconGeneral
BrandKonica Minolta
ModelKDR AU DDR
CategoryMedical Equipment
LanguageEnglish

Related product manuals