125
KMHA 500-000099
Revision D
KDR
TM
AU-DDR Advanced U-Arm System with
Dynamic Digital Radiography (KDR AU DDR)
VI. CLEANING
27.0 CLEANING MAINTENANCE
The KDR AU DDR has no user replaceable parts. Replacement of parts and/or preventative maintenance must be performed according to the KDR AU DDR Pre-In-
stallation, Installation, and Service Manual by a qualied eld service technician.
Table 27.1
Cleaning and Disinfecon Guidelines
Description Maintenance Interval
Reference
Document Page
Preamplier for Measuring
Chamber
Wipe with a damp cloth. As needed
Claymount Preamplier for
Measuring Chamber
6
KDR
Clean the exterior with a
soft lint-free cloth damp-
ened with a small amount
of anhydrous ethanol and
wrung well.
As needed
Konica Minolta KDR Sys-
tem Operation Manual
50
X-Ray Collimator
Clean varnished and alumi-
num surfaces with a damp
cloth, dry with a soft cloth.
Clean chrome surfaces
with a soft dry cloth only.
As needed
Series R 225 DHHS Manual
X-Ray Collimator
64
MTR 225 DHHS Manual
Radiological Collimator
65
Generator
(CPI)
Clean and re-grease the
HV connections on the HV
module.
Every six months AND
whenever a related certi-
able X-ray component is
replaced.
CPI CMP 200 DR Series
X-Ray Generator
6-1
Generator
(CPI)
Clean console surfaces
with soap and water. Disin-
fect console surfaces with
Antichlor or an equivalent
hypochlorite concentration
of 3-5%.
Every six months AND
whenever a related certi-
able X-ray component is
replaced.
CPI CMP 200 DR Series
X-Ray Generator
6-5
Generator
(Sedecal)
Clean external surfaces
with a soft cloth moistened
in clean water
As needed
Sedecal Maintenance, HF
Series Generators, Techni-
cal Publication MA-1004R6
6-7