142
KMHA 500-000099
Revision D
KDR
TM
AU-DDR Advanced U-Arm System with
Dynamic Digital Radiography (KDR AU DDR)
XI. APPENDIX B: MONTHLY CHECKLIST
35.0 MONTHLY CHECKLIST
Print these pages for use as a monthly checklist.
If there is any discrepancy use the table to make a note.
Hospital: ……………………………………………………………………………………………………………………………………………….
ID #: ……………………………………………………………………
Room: …………………………………………………………………
Sign: ………………………………………………………………….. Date: …………………………………………………………………..
1. Check the emergency stop. By acvang the emergency stop all motorized movements of the posioner are
prevented.
2. Measure between the X-ray tube focal spot and the acve image receptor surface of the image receptor holder.
The measured SID must correspond with the displayed SID. Check that the SID, shown on the display of both the
Image system and the collimator, correspond with the measured SID/FFD. The SID may dier by ±1%.
3. Check all hoses for damage.
4. Check all outer cabling for damage.
5. Clean all outer surfaces, except for the lubricated column segments. See the “Cleaning” chapter (Secon 34.0)
for cleaning instrucons.
6. Power up the stand and check all funcons.
7. Make movements with the stand and observe any irregularies.