6600 Product manual
Operator Manual
2
1.2 Intended use and contraindications
The system is intended use for medical education, scientific research, medical units
for head, abdomen, chest and limbs, bones of chest position and the table position, etc.
x radiographs for the diagnosis.
Contraindications: It is strictly prohibited within three months of pregnancy for
pregnant women to use the device to be checked.
1.2.1 Security classification
The Digital X-ray Radiography System covered by this Operation Manual is classified as:
Protection against electric shock: Class I -- Type B applied parts.
Mode of Operation: Continuous operation with intermittent loading.
Protection against harmful ingress of water: Ordinary.
Degree of safety in the presence of flammable anesthetics mixture with air or with
oxygen or with nitrous oxide: Not suitable for use in the presence of Flammable Anesthetics
Mixture with air or with oxygen or with nitrous oxide.
1.2.2 Product identification
The major items in the equipment have some identification labels attached to them
which provide the following manufacturer and product information.
Product.
Model.
Power condition (V), Frequency (Hz), Input power(kVA) Output power(kW)
Focal spot Anode kHU
Mode of operation
Date of manufacture.
Serial number.
Manufacturer.
Certification.
Label pasting instructions:
The unit label: Paste the label along the back of the column.
The part label: Paste the label on the obvious position of the corresponding
component, which is easy to identify.
1.3 Specification
1.3.1 Power condition
A) Power supply voltage and phase number: 3N~ 380V
B) Power frequency: 50Hz
C) The power resistance: 0.15 Ω or less
D) Power capacity: ≥60kVA