Chapter 8 System maintenance instruction
In order to ensure that the X-ray machine can continue to operate safely, it should be
regularly maintained. Maintenance has two aspects, one is maintenance by the user or the
operator, and the other is maintained by qualified X-ray maintenance personnel. The
maintenance procedures performed by the user or the operator are described below.
CAUTION
The professional maintenance and maintenance of this machine must be carried
out by professional maintenance personnel. The daily maintenance can be carried
out by the user.
User maintenance does not allow the generator cover to be removed. It is not
allowed to remove the internal components on the generator cabinet and console.
The manufacturer is not responsible for any injury to the personnel caused by the
violation of the regulations and damage to the machine.
Do not clean any part of the generator while the device is working. Turn off the
machine and turn off the power before cleaning.
Do not clean any part of the generator while the device is working. Turn off the
machine and turn off the power before cleaning.
The equipment room should be kept clean, dry, ventilated and ventilated. The
machine should avoid high temperature and direct sunlight.
Dehydration of filler in the high-pressure connector. Vaseline should be replaced
regularly. In the summer or during periods of heavy workload, it should be NOTEd
that regular inspections should be updated in a timely manner.
CAUTIONIn order to ensure that the product is in a good workpiece state,
during the use of the image workstation, always follow the instructions below
for inspection and maintenance, and record each inspection and maintenance
in the maintenance record.
Check the wrong Win XP error log. Clear the log file after the analysis is complete
Check the error log. Clear the log file after the analysis is complete
Scan the primary disk drive volume error using the Windows system utility.

Table of Contents