6600 Product Manual
Operator Manual
21
Requiements before installation
1)Standard toolkit for service engineers (Including M4 Allen key, cross and flat screwdrivers)
2)The high voltage generator must be installed in a system with protective functions, such
as overcurrent protection and disconnection of the system, and must be connected to the
necessary auxiliary equipment.
High voltage generator has minimum installation space requirement. The installation
position of the high voltage generator in the film room should be easy to maintain. The
following guidelines will contribute to the installation and maintenance of high voltage
generators.
1)If the high-voltage generator is fixed on the ground or floor, there should be a minimum
clearance space; 25 mm on each side of the minimum space is used for ventilation.
2)Please put the high-voltage generator in a well-ventilated environment. If the generator
is placed in an airtight cabinet, it should be considered to equip the generator with a fan. If
you have more requests, you can consult with our technicians.