instructs the 9400A to return a character string containing the
interval between data points used by a waveform processing function.
This value may be used as the <intval> parameter in the readout of
such waveforms if the user wishes to read only the computed data
points and none of the re-interpolated points.
3) REDEFINE (RDF) , ?
Instructs the 9400A
SELECTed trace,
to report the current
configuration of the
or
REDEFINE (RDF) , AVERAGE (AVG) , SUMMED , <maxpts>
<source> , <maxswps> , <reject> , <dither>
configures the SELECTed trace for a summed average,
or
REDEFINE (RDF) , AVERAGE(AVG) , CONTINUOUS(CONT) , <maxpts>
<source> , <weight>
configures the SELECTed trace for a continuous average,
or
REDEFINE (RDF) , EXTREMA (EXTR) <e-type> , <maxpts> ,
<source> , <maxswps>
configures the SELECTed trace for an extrema accumulation,
or
REDEFINE (RDF) , ARITHMETIC (ARI) <a-type> , <maxpts> , *
<sourcel> , <source2> , <m-fact> , <a-const>
configures the SELECTed trace for waveform arithmetic on two
sources,
or
10-9