EasyManua.ls Logo

Lloyd LS1 - Loadcell Calibration

Lloyd LS1
144 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LS SERIES SERVICE MANUAL
ISSUE 4.0 FULL CONTROL CONSOLE DIAGNOSTICS 7
12.6 Loadcell Calibration
Note that the loadcell can ONLY be calibrated if an accurate reference load is available.
Ensure that the Safe-Line button is pressed so that the yellow lamp is not lit.
1 Press the key under LOADCELL to display the "Calibrate Loadcell" screen.
Calibrate Load Cell
Serial
Number Size Calibrate View
XXXX 1kN Or Verify Points
2 Press the key under CALIBRATE OR VERIFY to display the screen shown below: -
Serial Number XXXXX Size XkN
Measured Load XXXXX N
Clear Cal Cal Direction
Points Point Tension Up
3 Attach the calibration device to the machine ensuring that no load is applied to the
loadcell or calibration device.
4 Zero the load reading on the calibration device.
5 Press the Zero key on the control console to zero the machine load display.
6 Ensure that UP is shown as the Direction. If DOWN is displayed, press the key under
DIRECTION so that the display shows UP.
7 Ensure that CAL TENSION is displayed above the 3rd key. If CAL COMPRES'N is
displayed, press the 3rd key so that the display shows CAL TENSION
8 The crosshead speed is selected using the numerical keypad as shown in the table
below. Note that the number 0 will stop the crosshead and hold the load. Do NOT press
the (Stop) key because the load may drop when the load pulls the crosshead back.
KEY SPEED
1 10 mm/min
2 5 mm/min
3 2 mm/min
4 1 mm/min
5 0.5 mm/min
6 0.2 mm/min
7 0.1 mm/min
8 0.05 mm/min
9 0.01 mm/min
0 ACTIVE STOP

Table of Contents

Related product manuals