p/n 18603-001, Rev. C LTV
®
1200/1150 Ventilator Service Manual Page 8-131
2) Replace the 4 black pan-head screws (P/N 10918B) into the turbine manifold. Screws
should be torqued to 20 in-oz (0.14 Nm).
3) Check the blue rubber grommets (P/N 10630) and grounding clips (P/N 10752) on the
oxygen blender mounting (4 each) for wear and replace if necessary (LTV
®
1200 only).
4) For the LTV
®
1200: Replace the four (4) blender mounting screws (P/N 10176), O
2
blender cap, and four (4) flat metal washers (P/N 10594). The grounding clips should be
positioned over the inside lip of the grommets, aligned with the edge of the side panel
and oriented as shown in the illustrations. Screws should be torqued to 60 in-oz (0.42
Nm).
For the LTV
®
1150: Replace the four (4) O
2
Block mounting screws (P/N 10627), Green
O
2
port cap (P/N 14446), and four (4) plastic shoulder washers (P/N 10630). Screws
should be torqued to 60 in-oz (0.42 Nm).