Selecta
PBL200051, Rev J
12 Safety and Regulatory
section 4.6.2 and Appendix C. For users who must comply with EN 207, the
safety eyewear must have a protection class of L5.
In addition to providing the required laser safety eyewear, take the following
steps to secure the treatment room, or the controlled area:
1 To alert personnel before they enter the controlled area, place a warning sign
on the outside of the treatment room door when the laser is in use.
2 Close the treatment room door during operation of the laser.
3 External door interlocks that automatically disable the laser when the
treatment room door is opened may be installed.
A blocking barrier, screen, or curtain capable of blocking or filtering the laser
beam could be placed to create a controlled area inside a large treatment
room. The barrier should be made of material that can withstand the power of
the treatment beam for the maximum exposure time, relative to the
configuration of the controlled area and the treatment parameters for the
specific medical application.