EasyManua.ls Logo

LUMENIS SELECTA - Electrical Requirements; Physical Characteristics

LUMENIS SELECTA
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications 93
Selecta
PBL200051, Rev J
Selecta Duet 532 nm laser Selecta Duet 1064 nm laser
Laser Safety Eyewear
Laser safety glasses with an OD5 @532
nm. Refer to the Laser Safety Eyewear sec-
tion of this manual for information.
Laser safety glasses with an OD5 @1064
nm. Refer to the Laser Safety Eyewear sec-
tion of this manual for information.
Electrical Requirements
100 -240 VAC, 50/60 Hz, 3.15 Amps 100 -240 VAC, 50/60 Hz, 3.15 Amps
Cooling
Air-cooled Air-cooled
Physical characteristics
Table height
Wheelchair accessible, 945 mm max, 695 mm min, travel ~250 mm
(37.8 in max, 27.4 in min, travel ~10.4 in)
Ta b l e w i d t h
Small: 630 mm (24.8 in) Medium: 880 mm (34.5 in)
Ta b l e d ep t h
Small: 400 mm (15.7 in) Medium: 450 mm (17.8 in)
Table wheel base
Small: 520 mm x 440 mm (20.4 in x 17.2 in)
Medium: 460 mm x 830 mm (18.1 in x 32.7 in)
Total system weight
<60 kg <60 kg
Environmental requirements (operating)
Temperature range
15°C to 40°C (59°F to 104°F)
Humidity
0 to 75% at 40°C, non-condensing
Altitude
up to 3,900 m
Environmental requirements (non-operating)
Temperature range
-10°C to 55°C (14°F to 131°F)
Relative humidity
0 to 90% at 55°C, non-condensing
Altitude
Standard commercial shipping altitude
Vibration
Capable of surviving transport by normal commercial recognized air, sea,
and land carriers
Shock
Lumenis shock requirements, based on size and weight