Preoperative Setup of the Slit Lamp and Laser 69
Selecta
PBL200051, Rev J
Verify that the slit spot is centered
1 Verify that the laser system is in standby mode.
2 Set the slit spot to 5 mm.
3 Set the microscope magnification to the maximum setting.
4 Visually verify that the slit spot is centered in the field of view. If the spot is
not centered, contact your Lumenis service representative.
Verify that the slit spot is centered
5 mm illuminated spot
Field of view