EasyManua.ls Logo

LUMENIS SELECTA - User Maintenance Procedures

LUMENIS SELECTA
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Selecta
PBL200051, Rev J
82 Maintenance
User Maintenance
Annual laser maintenance
Preventative maintenance, safety, power, and calibration checks should be
performed annually by a Lumenis-certified service engineer to ensure proper
laser performance.
Laser repair
All laser repairs should be performed by a Lumenis-certified service engineer.
For training and information, contact your local Lumenis service
representative.
Inspecting the laser system components
Before and after each use, inspect the laser system components for evidence of
dirt, debris, or damage.
Cleaning the slit lamp and laser module optics
Before each use, inspect the slit lamp oculars and objective lens for dirt, debris,
or damage. If necessary, clean the surfaces as follows:
1 Turn off the system power.
2 Wrap a piece of lens tissue (Kodak
®
or equivalent) around one end of a
cotton-tipped applicator (Puritan
®
or equivalent non-glued tip applicator).
3 Place several drops of reagent or 100% ethanol or methanol on the tissue.
Do not use ethanol or methanol to clean the external surfaces of the console
or slit lamp. Doing so may harm the surface.
4 Wipe the optic gently in one direction with lens tissue to remove all dust and
debris. Do not wipe in more than one direction, as loose particles might be
dragged across the surface and scratch the optical coating.
The optics have delicate coatings. When cleaning, wipe them gently, and avoid
getting fingerprints on them. If an optic appears to be damaged, contact your
local Lumenis representative.
Never use dry swabs or tissues to clean optics. They may damage the optical
coatings or scratch the surface.