Troubleshooting Guide 81
Selecta
PBL200051, Rev J
“Energy High > 50%”
or “C101” displays on
the remote control
The detected energy level was at least
50% higher than the level selected.
Place the laser in ready mode. If the
condition continues, restart the laser. If
the condition persists, contact your local
Lumenis representative.
“Energy Low < 50%” or
“C102” displays on the
remote control
The detected energy level was at least
50% lower than the level selected.
Place the laser in ready mode. If the condi-
tion continues, restart the laser. If the con-
dition persists, contact your local Lumenis
representative.
“Over Temperature”
or “W106” displays on
the remote control
The laser temperature is too high. Place the laser in standby mode. Allow the
laser to cool down. Place the laser in ready
mode to continue treatment. If the
condition continues, contact your local
Lumenis representative.
“Footswitch Not
Present” or “W108”
displays on the remote
control
The footswitch is not properly
connected to the laser.
Refer to the Operation section of this
manual, and connect the footswitch as
instructed. If the condition continues,
restart the system, ensuring that the
footswitch is not pressed during start up.
If the condition persists, the footswitch
may be defective; contact your local
Lumenis representative.
The footswitch is properly connected,
but the footswitch laser activation mode
is not selected on the remote control
options screen.
Refer to the Remote Control Options
Screen in the Operation section of this
manual for instructions on selecting the
footswitch laser activation mode.
“No LaserLink” or
“W110” displays on the
remote control.
The LaserLink S delivery device is not
properly connected.
Refer to the LaserLink connection
instructions in the LaserLink S operation
manual, and connect the LaserLink S S as
instructed.
If any other code
displays that is not
explained here...
The error condition is not user correct-
able.
Contact your local Lumenis
representative.