Selecta
PBL200051, Rev J
92 Maintenance
Specifications
The Selecta Ophthalmic laser system delivers both a 1064 nm and 532 nm
treatment beam; therefore, the laser specifications outlined below for each
wavelength are applicable to the Selecta Duet model.
Selecta Duet 532 nm laser Selecta Duet 1064 nm laser
Laser Specifications
Ty p e
Frequency-doubled, Q-switched Nd:YAG Q-switched Nd:YAG
Principal output—wavelength
532 nm wavelength 1064 nm wavelength
Operating mode
Frequency doubled, pulsed Fundamental, pulsed
Pulse duration
3 nanoseconds 3 nanoseconds
Pulse mode
single pulse single, double, or triple pulse
Energy
0.3 to 2 mJ 0.3 to 30 mJ
Maximum Rep Rate
3.0 Hz 3.0 Hz
Laser beam spot size
400 µm at visual focal plane 8 µm
Cone angle
<3° 14° to 16°
Posterior Offset
N/A 100 µm nominal, continuously adjustable
from 0 to 350 µm posterior, with detents
at 0, 100, 250, and 350 µm.
CDRH classification
Class IIIb Class IIIb
IEC 60825 classification
Class 3B Class 3B
Aiming Beam Specifications
Ty p e
CW Diode laser CW Diode laser
Power
300 µW maximum 150 µW maximum
Principal output
635 nm wavelength 635 nm wavelength
CDRH Classification
Class II Class II
IEC 60825 classification
Class 2 Class 2