PBL200051, Rev J
120 Appendix 1
Guidance
and
Manufacturer's
Declaration:
Electromagnetic
Immunity
Selecta ophthalmic laser family
is
intended
for
use
in
the
electromagnetic
environment
specified
below.
The
customer
or the
user
of
Selecta ophthalmic laser family
should
ensure
that
it
is
used
in
such
an
environment.
Immunity
Te s t I E C
60601
Te s t
Level
Compliance
Level
Electromagnetic
Environment: Guidance
Electrostatic
Discharge
(ESD)
IEC61000-4-2
±6kV
contact
±8kV
air
Class A
Floors
should
be wood,
con-
crete,
or ceramic
tile.
If
floors
are
covered
with synthetic
material, the
relative
humidity
should
be
at
least
30%.
Electrical
fast
transient/burst
IEC61000-4-4
±2
kV
for
power supply
lines
±1kV
for
input/output
lines
Class A
Mains
power
quality should
be
that
of
a typical
commercial
or
hospital environment.
Surge
IEC61000-4-5
±1kV
line(s) to line(s)
±2kV
line(s) to earth
Class A
Mains
power
quality should
be
that
of
a typical
commercial
or
hospital environment.
Vo l t a g e
dips,
short interrup-
tions
and
voltage variations
on
power
supply input
lines
IEC61000-4-11
<5%
U
T
(>95%
dip in
U
T
)
for
0.5
cycle
40%
U
T
(60%
dip
in
U
T
)
for
5
cycles
70%
U
T
(30%
dip
in
U
T
)
for
25
cycles
<5%
U
T
(>95%
dip in
U
T
)
for
5
sec.
Class A
Class A
Mains
power
quality should
be
that
of
a typical
commercial
or
hospital environment.
If
the user
of
Selecta oph-
thalmic laser family requires
continued operation
during
power
mains interruptions,
it
is recommended
that Selecta
ophthalmic laser family
be
powered from
an uninterrupt-
ible power
supply
or
a bat-
tery.
Power
frequency
(50/60Hz)
magnetic
field
IEC
61000-4-8
3
A/m
N/A
Power-frequency magnetic
fields should
be
at
levels char-
acteristic
of
a typical
location
in
a typical
commercial or
hospital environment.
NOTE:
U
T
is
the
a.c.
mains
voltage
prior
to
application
of