EasyManua.ls Logo

LUMENIS SELECTA - Page 79

LUMENIS SELECTA
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting Guide 79
Selecta
PBL200051, Rev J
Slit lamp optics are dirty. Inspect and clean the slit lamp optics, as
detailed in “User Maintenance” in this
chapter.
No treatment beam is
delivered when the
footswitch or joystick
pushbutton is pressed,
or the beam is of poor
quality
The laser is in standby mode.
Place the laser in ready mode.
The laser activation mode is not properly
selected on the remote control options
screen.
As detailed in the “Laser activation mode
section of this manual, verify that the cor-
rect laser activation mode is selected on
the remote control options screen. For
example, if you are using the footswitch to
activate the treatment beam, ensure that
the footswitch mode is selected.
The laser components are not properly
connected.
Refer to the Operation section of this
manual, and check the laser system con-
nections.
The laser module optics are dirty or mis-
aligned.
Contact your local Lumenis representa-
tive.
Inadequate laser power The laser module optics are dirty or mis-
aligned.
Contact your local Lumenis
representative.
Cannot select the
Ready mode
The LaserLink S is attached and the
Selecta system is in YAG or SLT laser
mode.
Remove the LaserLink S.
The slit lamp table is in motion. Adjust the table to the desired height and
position, and then select the Ready mode.
The incorrect illumination mirror is
installed.
Install the single illumination mirror, as
described in the LaserLink S operator
manual.
No or inadequate slit
lamp illumination
The illumination intensity control is not
in the proper position.
Refer to the Operation section of this
manual and adjust the illumination inten-
sity control.
The illumination bulb is burned out. Refer to “Changing the Slit Lamp Illumi-
nation Bulb” in the Maintenance section
of this manual for information on replac-
ing the halogen illumination lamp.
Symptom Probable Cause Suggestion

Table of Contents