EasyManua.ls Logo

Malvern Instruments Spraytec - Power Requirements; Power Cords

Default Icon
221 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Provide enough space to allow easy access to all components of the system. Allow
for 2500mm (width) x 850mm (depth) of bench space. Allow at least 800mm
above the bench surface for access to the cell area.
Power requirements
The mains power supply must be clean and filtered. If necessary, fit an
Uninterruptible Power Supply (UPS) to remove any spikes or noise. The power
requirement for each component of the Spraytec system is given below.
Component Power requirement
Power sockets
required
Spraytec ~ 100-240V, 50-60Hz 1
NSS ~ 100-240V, 50-60Hz 0
Switching box ~ 100-240V, 50-60Hz 1
Extractor ~ 100-240V, 50-60Hz 1
Computer See manufacturer’s documentation 1
Computer monitor See manufacturer’s documentation 1
Power cords
Power cord set requirements
Power cord sets must meet the requirements of the country where the product is
used. For further information on power cord set requirements, contact a Malvern
representative.
General requirements
The requirements listed below are applicable to all countries:
.
The power cord must be approved by an acceptable accredited agency re
-
sponsible for evaluation in the country where the power cord set will be in
-
stalled.
APPENDIX C