Continuous spray SOP
This section describes how to create an SOP for use with a continuous spray, for
example a coatings or agrochemical spray.
.
In the Welcome dialogue select the Default continuous template and
click Next.
.
The SOP Type dialogue is displayed:
Click the Hardware configuration button and select Open bench
as the hardware unless using extraction, in which case select Open bench
with extraction. Under Basic settings keep the measurement type as
Continuous and sample at one measurement a second or less. Click Next.
'
Note
.
The Lens type is automatically detected. If the lens specified here is not
installed on the instrument, the SOP will not run (see Chapter 8).
.
In the Labels dialogue specify any required information on the spray, its
source and lot, etc. to allow the spray to be traced in the future. If required,
the operator can change/add to these when they run the test. Click Next.
.
In the Instructions dialogue type any instructions for operators to follow
at the SOP start or end. For either, select the relevant check box and type in
the instruction text below it. These instructions will pop up on the screen
when an operator runs the SOP. Click Next.