Operating Instructions – Labmaster
©
130 and MB20G Systems
Instruction Manual
Chapter 10: Maintenance & Troubleshooting
© MBraun, Inc. Edition 1/2007 Page 10-9
Place an absorbent lint free cloth under the probe
and moisten the probe with distilled water.
Wrap the moistened cloth around the probe. Gently
and slowly spin the cloth around the probe to clean
it.
Dry the probe in the same manner with a dry lint
free cloth.
Do not clean the probe by moving the cloth in
an up and down motion. This will damage the
coil and will render the analyzer unusable
resulting in the need for repair by M.Braun.
Place an absorbent lint free cloth under the probe
and moisten the probe with Phosphoric Acid.