Medica EasyStat - 7 Specifications

Medica EasyStat
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
110
7. Specifications
CLIA Classification: Moderate complexity
Sample Type: Whole blood
Sample Size: 120 µL Syringe mode/95 µL Capillary mode
Measured Parameters and Limits*
PO
2
5 - 700 mmHg PCO
2
5.0 - 150.0 mmHg pH 6.500 - 8.000 pH units Hct 10-70%
Na
+
80-200 mmol/L K
+
1.0-20.0 mmol/L Ca
++
0.25-5.00 mmol/L (Ca
++
mode) Cl
-
50.0-150.0 (Cl
-
mode) mmol/L
Calculated Parameters
THb (Total Hemoglobin) 3.3-23.3 g/dL
pH (T) (pH temperature corrected)
PCO
2
(T) (PCO
2
temperature corrected)
PO
2
(T) (PO
2
temperature corrected)
TCO
2
(Total Carbon dioxide)
0-50 mmol/L
HCO
3
-
(Bicarbonate) 0-50 mmol/L
BE
b
(Base Excess in blood)
-25.0 to 25.0 mmol/L
BE
ecf
(Base Excess in extra cellular fluid) -25.0 to 25.0 mmol/L
SBC (Standard Bicarbonate) 0-50 mmol/L
%SO
2c
(Oxygen Saturation) 40.0 - 100.0% (calculated at normal P
50
)
CtO
2
(Oxygen Content) 3.0 - 30.0 mL/dL
A-aDO
2
(Alveolar arterial oxygen gradient) 0-700 mmHg
RI (Respiratory Index) 0.0 - 70.0
Ca
++
(7.4) (for 7.2 < pH < 7.6) (Ca
++
mode) 0.22 - 5.58 mmol/L
Anion Gap (Cl
-
mode) 10-75 mEq/L
Input Parameters
Patient Temperature (20 - 45°C) Time Drawn (00:00)
Hemoglobin (3.0 - 30.0 g/dL) Sample Source (arterial, mixed venous, venous)
FIO
2
(Fraction Inspired Oxygen) (10 - 100%) Sample Type (radial, brachial, femoral, arterial line, cord)
Patient ID (14 digits)
Operator ID (14 digits)
Sample Temperature Control: 37.0°C ± 0.2°C
Environmental Conditions: Indoor Use
15–30°C (59-86°F), 500–800 mmHg (max 15 PSI)
Altitude up to 2000m
5–85% relative humidity, non-condensing
atmospheric air environment (21% O
2
)
Overvoltage category II, Pollution Degree 2
Analysis Time: <120 seconds
Data Storage: 64 Patient results with Operator ID, Patient ID, Date and Time
QC—up to 93 results for each Level (Blood Gas/Electrolytes 1, 2, 3, Hct 1, 2)
Calibration: Automatic or On-Demand
Input/Output:
Numeric keypad, graphic display, 27 column thermal line printer,
barcode reader port, RS-232 computer interface port
Power: 100/115~VAC, 50-60 Hz, 0.8 A or 220~VAC, 50-60 Hz, 0.4 A
Refer to the chassis serial number label for the voltage that has
been factory set on your analyzer, and for proper fuse replacement.
Size & Weight: 14.5"W x 12.5"H x 7.0"D (37cmW x 32cmH x 18cmD), 17lbs (7.7kg) with reagent module
*Limits are using default correlations. Refer to CORRELATION for more information.