EasyManua.ls Logo

Medica EasyStat - Air Messages for One Reagent; Air Messages for Two or more Reagents

Medica EasyStat
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
127127
t r o u b l e s h o o t i n g
g
AIR IN CALIBRANT A AIR IN CALIBRANT B
AIR IN CALIBRANT C AIR IN SAMPLE
AIR IN CONTROL AIR IN CLEANER
FLOW
Air messages for one reagent (Calibrant A, Calibrant B
or Rinse)
Perform the REAGENT FLOW test under TEST
FLUIDICS in DIAGNOSTICS with the sampler
in the closed position. If all reagents pass, recalibrate.
If any reagent fails, remove then reinstall the reagent
module.
Repeat the REAGENT FLOW test.
If any reagent fails, remove the valve mod-
ule and flush with water as described under
VALVE MODULE/VALVE ERROR. Repeat the
REGAGENT FLOW test (sampler closed).
If failure continues, install a new reagent module.
Air messages for two or more reagents (Calibrant A,
Calibrant B and Calibrant C)
Perform the SAMPLE FLOW test under TEST
FLUIDICS in DIAGNOSTICS.
Fill the troubleshooting kit syringe with red test dye solu-
tion.
Lift the sampler into the syringe mode and use the red test
dye as a sample.

Table of Contents