o p e r a t i n g t h e a n a l y z e r
r
BAROMETRIC PRESSURE
Select BAROMETRIC PRESSURE to view the barometric pres-
sure. When the displayed barometric pressure does not match a ref-
erence barometer, set the analyzer’s barometric pressure by proceed-
ing to the SETUP MENU and selecting USER OPTIONS,
then BAROMETRIC PRESSURE.
Caution: The barometer is calibrated at the factory. The barometric
pressure setting should only be changed to correlate it to a local refer-
ence barometer. The proper operation of the analyzer depends on a
correct barometric pressure setting.
TEMPERATURE
Select TEMPERATURE to view the sensor, valve and room tem-
peratures. The readings must be between:
Sensor 36.9° to 37.1°C
Valve 36.0° to 41.0°C
Room 15° to 30°C
If any temperature is not within range, refer to the Troubleshooting
chapter in this manual.
PRIME FLUIDS
Prime the fluid path by pumping Calibrant B, Calibrant A and
Calibrant C from the reagent module.
PRINT mV
,
S
Select PRINT mV
,
S to display and print millivolt readings for
each parameter. This feature is useful when reporting problems to your
EasyStat distributor.