EasyManua.ls Logo

Medica EasyStat - Sample Probe;Sensor Module Obstructions

Medica EasyStat
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
131131
t r o u b l e s h o o t i n g
g
s
Sample probe/sensor module obstructions
To remove a persistent flow obstruction, or as preventive maintenance,
perform the following steps.
Remove the electrodes from the sensor module. Refer
to Replace Components/Electrodes in Operating the
Analyzer.
Install the test block from the troubleshooting kit into the
sensor module. Release the compression plate.
Disconnect the pump tubing from the bottom of the sensor
module.
Lift the sampler to the syringe mode position.
Attach the large tubing segment to the syringe from the
troubleshooting kit. Fill the syringe with a 10% bleach
(0.4-0.6% NaClO) solution.
Connect the tubing segment to the bottom of the sensor
module.
Hold a container below the sampler.
Flush the flow path by gently pushing on the syringe.
Allow the bleach to remain in the flow path for 2 minutes.

Table of Contents