EasyManua.ls Logo

Medica EasyStat - Pump Cal Failures; Flow

Medica EasyStat
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
125125
t r o u b l e s h o o t i n g
g
FLOW
To remove or replace the EasyStat sensor module, pump tubing or
valve module, follow the instructions under Replace Components in
Operating the Analyzer.
s
PUMP CAL FAIL BUBBLE DETECTOR!
Pump Cal failures (PUMP CAL FAIL)
Verify that all EasyStat components (electrodes, valve
module, pump tubing and reagent module) are installed
correctly.
Confirm that the sampler is down.
Open the access door.
Perform the
PUMP CAL
test under
TEST
FLUIDICS
in the DIAGNOSTICS menu. Look for
liquid to flow through the pump tubing.
If PUMP CAL fails, but flow is present, the pump tub-
ing may be defective. Install a new pump tube and retest.
If the error persists, replace the probe wiper.
If flow is not present during the PUMP CAL test,
proceed to the troubleshooting instructions for FLOW/
AIR” messages.

Table of Contents