o p e r a t i n g t h e a n a l y z e r
S
li
g
htly Lift
S
am
p
ler
Sensor Module/Sampler
Sensor module or sampler replacement should be infrequent. Sensor
Module replacement will be necessary only if certain display mes-
sages become chronic (refer to SENSOR MODULE in Troubleshooting).
Sampler replacement is necessary if the sample probe is bent, dam-
aged or obstructed (refer to FLOW in Troubleshooting), or if the sam-
pler movement is defective.
Select SENSOR MODULE/SAMPLER. Fluid is automatically
purged from the sample flow path. When the purge cycle is complete,
proceed with sensor module and/or sampler replacement.
To remove the sensor module from the analyzer, first remove the pump
tubing.
Disengage the sampler by slightly lifting the sampler until it unlocks
from the valve module.