39
o p e r a t i n g t h e a n a l y z e r
r
s
s
Users should follow applicable agency requirements for quality control
testing.
Recommended Material
EasyQC Quality Controls:
Level 1
Level 2
Level 3
Hct Level 1
Hct Level 2
Do not use perfluorocarbon-based control materials on the EasyStat
analyzer. This type of control material will damage the electrodes.
Precautions
Refer to the package insert provided with the quality control material.
Recommended Procedure
Run all control levels each day the analyzer is in use.
Data is stored for
future statistical analysis.
Use Instructions
Refer to the package insert provided with the quality control material.
Storage and Stability
Refer to the package insert provided with the quality control material.
Handling
Refer to the package insert provided with the quality control material.
Expected Results
Refer to the package insert provided with the quality control material.