91
o p e r a t i n g t h e a n a l y z e r
r
CRITICAL
Set the critical ranges for each parameter. Select arterial, mixed
venous, or venous. Press the number keys to enter the desired lim-
its, and YES to confirm each individual setting and move to the next
location. When all limits are entered, press YES to CONFIRM?
to accept all entries. When NO is pressed as a response to
CONFIRM?, entries can be corrected.
When analysis results are in the critical range, they are flagged
as low (↓↓) or high (↑↑) on both the display and the printout. For
example, when the critical pH limits are 7.300 and 7.600, a ↓↓ flag
is displayed/printed for patient pH results less than or equal to 7.300
or a ↑↑ flag is displayed/printed for patient pH results greater than
or equal to 7.600.
Patient results that are outside the normal range, and not within the
critical range will be flagged as low () or high () on the display
and printout.
RILIBAK
Set the RiliBAK target, percent deviation and range for each param-
eter. To change the setting, press NO to CONFIRM?. Press the
number keys to enter the desired limit and press YES to CONFIRM?
to accept all entries. When NO is pressed as a response to
CONFIRM?, entries can be corrected.
AMR
Set the Adjustable Measurement Range for each parameter. To
change the setting, press NO to CONFIRM?. Press the number keys
to enter the desired limit and press YES to CONFIRM? to accept
all entries. When NO is pressed as a response to CONFIRM?,
entries can be corrected.
Na
+
115
.
0
--
150
.
0
mmol/L
Hct
33
--
70
%
SET
REFERENCE
LIMITS
CRITICAL
ARTERIAL
pH
7
.
300
--
7
.
600
PCO
2
20
.
0
--
50
.
0
mmH
g
PO
2
60
--
250
mmH
g
CONFIRM?
SET
REFERENCE
LIMITS
CRITICAL
ARTERIAL
K
+
Ca
++
0
.
82
--
3
.
00
mmol/L
CONFIRM?
REFERENCE LIMITS 1/6
Rili-BAK LEVEL 1
LOT #XXXXXXX
pH
TARGET X.XXX mmHg
DEVIATION X.XX %
RILIBAK X.XXX
-
X.XXX mmHg
sCa
++
is replaced with Cl
-
when Cl
-
mode is selected.