EasyManua.ls Logo

Medica EasyStat - 4 Operating the Analyzer; Calibrate

Medica EasyStat
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
Calibrate
3
LIFT
SAMPLER
TO
ANALYZE
2
2
Hct
Na
+
Ca
++
CALIBRA
TIO
A
A
N
HOME
MENU
1
*
2
*
3
CALIBRATE
4
DAILY
CLEANER
5
SECOND
MENU
HOME
MENU
1
ANALYZE
SAMPLE
2
ANALYZE
QC
3
CALIBRATE
4
DAILY
CLEANER
5
SECOND
MENU
SLOPES
pH
61
.
02
PCO
2
53
.
17
P
O
2
3.0
1
Hct
18
.
88
Na
+
58
.
98
K
+
61
.
50
Ca
++
24
.
62
CALIBRATION
BUBBLE
DETECTOR
CAL
air
fluid
BD
PASS
171
030
PUMP
CAL
PUMP
PASS
0799
REAGENT
MODULE
STATUS
021%,
15
DAYS
JUNE
-
05
-
20
;13
:
35
4. Operating the Analyzer
s
The EasyStat analyzer requires a two-point calibration after instal-
lation and warm-up. Automatic two-point calibrations occur at
pre-selected intervals chosen under CAL FREQUENCY in the
CONFIGURATION section of the SETUP MENU. Automatic
Hct calibrations are included in each two-point calibration. On-
demand two-point calibrations can be initiated in the HOME MENU
by choosing CALIBRATE. If not calibrated, the HOME MENU
only displays options 3, 4 and 5.
The ACCESS DOOR must be closed during warm-up, calibra-
tion, and sampling.
Select CALIBRATE to perform a two-point calibration and prepare
the EasyStat analyzer for sample analysis. The analyzer displays the
calibration status. Bubble detector and pump cal results are printed
at completion. After a successful calibration, the analyzer displays
and prints the slope of each electrode, the % of calibrants remaining,
and the number of days before expiration. After completion of a suc-
cessful calibration, the analyzer displays LIFT SAMPLER TO
ANALYZE pH PCO
2
PO
2
Hct Na
+
K
+
Ca
++
.
Ca
++
is replaced with Cl
-
when Cl
-
mode is selected.

Table of Contents