87
o p e r a t i n g t h e a n a l y z e r
r
sCa
++
is replaced with Cl
-
when Cl
-
mode is selected.
CONFIGURATION
Select CONFIGURATION to set analyzer options. Select from:
PO
2
/PCO
2
/BP, Na
+
/K
+
/Ca
++
, Hct, Hb, CAL
FREQUENCY, AUTO STANDBY, and TEMPERATURE.
PO
2
/PCO
2
/BP
Change the measurement units (select mmHg or SI units) for PO
2
,
PCO
2
and barometric pressure. Results for SI units are shown in KPa.
Na
+
/K
+
/Ca
++
or Cl
-
Change the measurement units (select mmol/L or mmEq/L units).
HCT
Change the measurement units (select % or V/V) for hematocrit.
Hb
Change the measurement units for Hb. Select mmol/L, g/L or g/dL.
BC Checksum
Turn the feature ON or OFF. When the feature is ON, the analyzer
has the uses the last digit in the barcode to perform a barcode check-
sum calculation.
CAL FREQUENCY
Press the number keys to select and change an option. Press YES to
CONFIRM? all configuration settings.
Select the time interval between automatic two-point calibrations.
Select from: 2 hours, 4 hours, 8 hours
1 PO
2
/ PCO
2
/ BP mmHg
2 Na
+
/
K
+
/ Ca
++
mmol/L
3 Hct %
4 Hb
g/dL
5 BC CHECKSUM OFF
CONFIRM ?
CONFIGURATION 1/2
1 CAL FREQUENCY 8 HRS
2 AUTO STANDBY OFF
3 TEMPERATURE C
4 Rili-BAK OFF
5 AMR OFF
CONFIRM ?
CONFIGURATION 2/2