o p e r a t i n g t h e a n a l y z e r
s
AUTO STANDBY
Select ON or OFF. When ON, the analyzer will enter standby auto-
matically. Standby is available manually with either selection.
TEMPERATURE
Change the temperature unit. Select Fahrenheit (F) or Celsius (C).
RILIBAK
Turn RiliBAK ON or OFF to use the statistic-based method for perform-
ing QC analysis required in Germany.
AMR
Turn AMR (Adjustable Measurement Range) ON or OFF. When ON,
the default values are the low and high instrument range for each
parameter. The ranges can be changed to ranges within the default
limits.
PRINTER OPTIONS
Select printing options.
PRINTER
Turn the printer ON or OFF. When the analyzer runs out of printer
paper, turn the printer OFF to prevent damage to the mechanism.
CALC D VALUES
Turn printing of calculated values ON or OFF.
PRINTOUTS
Select up to four (4) sample printouts.
SET REFERENCE LIMITS
Select SET REFERENCE LIMITS to set the quality control limits
for each level, and the normal and critical patient limits.
It is recommended that any stored quality control results be deleted
prior to entering information for a new lot of quality control. Refer to
DELETE DATA in the SETUP MENU.