EasyManua.ls Logo

Medica EasyStat - Page 138

Medica EasyStat
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
132132
t r o u b l e s h o o t i n g
Air was detected in the sample.
Confirm that sample volume is sufficient (syringe mode -
120 µL, capillary mode - 95 µL).
There may be air bubbles in the sample itself (sample was
shaken). Anlyze a sample without air bubbles.
If the error continues, refer to FLOW in this chapter.
These messages indicate a small movement of cali-
brant, patient sample, or quality control material dur-
ing the measurement period of a calibration or analysis
cycle.
When any of these messages appear, repeat the calibra-
tion or sample analysis to confirm reproducibility.
Continued MOVEMENT messages could indicate damaged elec-
trode O-ring seals, or a partial obstruction in the sample flow path.
Obstructions are most common in the sample probe, sensor module, or
electrodes. Follow the instructions for obstruction removal described ear-
lier in this section under FLOW/Sample probe/sensor module obstruc-
tions.
Remove the syringe and tubing from the sensor module.
Discard any unused bleach from the syringe and fill the
syringe with water.
Flush the flow path with water. Repeat 4 times using a
syringe full of water each time.
Install the sensor module and reconnect the pump tubing.
Confirm a successful SAMPLE FLOW test with the test
block installed.
If an obstruction remains, remove the sampler from the
sensor module (refer to Replace Components/Sensor
Module in Operating the Analyzer)
Take a syringe filled with water and attach with small tub-
ing segment to the sampler in the syringe mode and flush
the flow path.
Install a new sampler if the obstruction cannot be removed
Confirm a successful SAMPLE FLOW test with the test
block installed. Remove the test block and reinstall the
EasyStat electrodes.
Release the compression plate and close the sampler.
Perform the SAMPLE FLOW test. If successful, resume
normal operation.
If the failure continues, it is the bubble detector that is
obstructed. Placing a warm damp paper towel at its outlet
for a period of time usually loosens the material block-
ing the flow path and it can then be flushed. Perform the
SAMPLE FLOW test. If successful, resume normal operation.
If an obstruction remains, replace the sensor module.

Table of Contents