EasyManua.ls Logo

Medica EasyStat - Reagent Module Error Handling

Medica EasyStat
158 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
128128
t r o u b l e s h o o t i n g
Look for red dye movement through the pump tubing. If
the red test dye does not move, proceed to the instructions
for “No red dye movement during FLOW test”.
If there is fluid movement, this indicates a poor sampler/
valve module seal.
Inspect the sampler/valve module interface.
Move the sampler to the capillary position.
Inspect the seal on the sampler and the top of the valve
module for debris. Clean these areas with a 10% bleach
(0.4-0.6% NaClO) solution on a damp cloth, then repeat
the SAMPLE FLOW test.
If the error persists, replace the probe wiper and retest.
If failure continues, the valve module or the reagent mod-
ule may be defective.
Reinstall the reagent module and repeat the REAGENT
FLOW test (sampler closed).
If successful, resume operation.
If failure continues, remove the valve module and flush
with water as described under VALVE MODULE/VALVE
ERROR. Repeat the REAGENT FLOW test (sampler
closed).
If failure continues, install a new reagent module.

Table of Contents