o p e r a t i n g t h e a n a l y z e r
r
s
The analyzer displays CALCULATED VALUES?. The calcu-
lated results are printed with the measured results. View the calculated
values by pressing YES. Calculated values for THb, pH(T), PCO
2
(T),
PO
2
(T), TCO
2
, HCO
3
-
, and BE
b
are displayed. MORE DATA?
appears at the top of the display screen. Press YES to view the cal-
culated results for BE
ecf
, SBC, %SO
2
c, O
2
ct, A-aDO
2
, RI, and Ca
++
(7.4). In Cl
-
mode, Ca
++
(7.4) is replaced with Anion Gap.The calcu-
lated results use the following default values, if patient values were not
entered (see PATIENT INFORMATION screen in the SETUP
MENU):
Patient Temp 37°C
Hb 14.5 g/dL
FIO
2
21%
User entered correlation values are applied to the measured results to
provide the displayed and printed results. If correlations are applied,
reported results will be printed as CORRELATION APPLIED
results. Refer to CORRELATION under USER OPTIONS in
the SETUP MENU for more information.