EasyManua.ls Logo

Microgate Optogate - Jump Test

Default Icon
170 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Version 1.9 Page 62 of 170
The available fields for the different test types are the following:
4.2.2.2 JUMP TEST
Main Parameters
Start type: indicates if the start of the test must take place with a ‘status change’ (i.e.
entering or exiting the bars) or with an ‘external impulse’ coming from the jack sockets (start
button, signal light, etc.).
Start: indicates if at the beginning of the test the patient must be inside the area’ of the bars
or ‘outside the area’.
Stop Type: as for the start, the end of the test can take place because of a Status change’ or
an external impulse’, or (in the case of time tests) the end of the test time (e.g. a test “Jumps
15 seconds” will have the Stop type = ‘End of Time’).
Finish: for the Stop type = ‘Status change’ it is necessary to specify if, at the end of the test,
the patient must be inside the area’ of the bars or outside the area’; in the other two Stop
type cases this field does not appear.
Number of Jumps: appears only if the Stop type = Status changeand specifies how many
times the patient must jump (1 to 99).
Test Length: appears only if Stop type = End of Time and specifies how many
minutes:seconds the test must last (from 00:01 to 09:59).
Secondary Parameters
Minimum contact time [ms]: allows you to set a minimum contact time in milliseconds; if
the system registers a contact time shorter than this value, it is considered incorrect and
added to the associated flight time (usually the previous one). This control can be
enabled/disabled by inserting the value 0.
Minimum flight time [ms]: allows you to set a minimum flight time in milliseconds; if the
system registers a flight time shorter than this value, it is considered incorrect and added to
the associated contact time. This control can be enabled/disabled by inserting the value 0.
Maximum flight time [ms]: allows you to set a maximum flight time in milliseconds; if the
flight time is higher, it is considered incorrect; if the system registers a flight time longer than
this value, the time is not considered.
Starting foot: if known, the foot with which the test starts (“rightor “left”) can be indicated,
otherwise leave as “not defined”.

Table of Contents

Other manuals for Microgate Optogate