EasyManua.ls Logo

Microgate Optogate - Tapping Test

Default Icon
170 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Version 1.9 Page 69 of 170
4.2.2.7 TAPPING TEST
Main Parameters
Start type: indicates if the start of the test must take place with a status change (i.e.
entering or exiting the bars) or with an ‘external impulse’ coming from the jack sockets
Start: can only be ‘inside the area
Staring foot: if known, the foot with which the test starts (“right” or left”) can be indicated,
otherwise leave as “not defined
Stop type: indicates how the end of the test must be interpreted, (i.e., as ‘External impulse
like a photocell at the end of the lane) or as ‘Timeout
Test Length: appears only if Stop type = End of Time and specifies how many
minutes:seconds the test must last (from 00:01 to 09:59)
Secondary Parameters
Minimum contact time [ms]: see above
Minimum flight time [ms]: see above
Maximum flight time [ms]: see above
Minimum gap between feet [cm]: see above
Minimum foot length [cm]: see above
In order to perform a Tapping Test, it is important that the patient’s feet are vertical to the OptoGait
bars.
Figure 53 Correct execution of a Tapping Test

Table of Contents

Other manuals for Microgate Optogate