507- The edit session for <name> must be saved before it is prepared for
analysis. Save changes and proceed with the analysis? (yes, no)
Cause: You are trying to start a degas operation using a file that con-
tains unsaved changes and is open for editing.
Action: Select Ye s to save the changes and proceed with the degas op-
eration; select No to cancel the degas operation and continue ed-
iting the Sample Information file.
508 Cannot open sample information file (name). It is already in use.
Cause: You tried to select a file for automatic degassing that had al-
ready been selected.
Action: Choose a file that is not already selected for automatic
degassing.
Error Messages SmartPrep
B-10 Aug 98