Specifications
The SmartPrep degasser has been designed and tested to meet the specifications
provided in the following table.
Table 1-1. SmartPrep Specifications
Characteristic Specification
Degas Method Flowing gas
Number of Stations Six
Temperature Range Ambient to 425
o
C, in 1
o
C increments; each
station is controlled separately
Accuracy
±5
o
C of set point for temperatures up to 100
o
C;
±10
o
C for temperaures between 100 and 300
o
C;
±15
o
C for temperatures over 300
o
C
Soak Time 0 to 999 minutes, in 1-minute increments
Ramp and Soak Cycles Up to 5 ramp and soak cycles
Ramp Rate From 5
o
C/min. to 20
o
C/min.
Electrical
Volt age
Power
Frequency
85 to 265 VAC
500 VA Maximum
50/60 Hz
Environment
Temperature
Humidity
10 to 35
o
C, operating; 0 to 50
o
C, nonoperating
20 to 80% relative, noncondensing
Gases Used Nitrogen and other inert gases
Flow Rate 0 - 500 cm
3
/min.
Physical
Height
Width
Depth
Weight
50 cm (20 in.)
50 cm (20 in.)
40 cm (16 in.)
15 kg (33 lbs)
SmartPrep Specifications