Figure 2-3. Cleaning Sample Tubes
5. Using rubber gloves, remove the sample tubes and filler rods from the
reservoir.
6. Clean the interior of the sample tubes with a brush.
7. Rinse the sample tubes and filler rods thoroughly with hot water. Then rinse
them with isopropyl alcohol or acetone using a waste container to collect
used solvent.
If isopropyl alcohol or acetone is not available, deionized water may be
used to rinse the sample tubes.
Figure 2-4. Rinsing Sample Tubes
SmartPrep Preparing Samples for Degassing
Sep 98 2-11