PRODUCT INFORMATION
Milli-Q Synthesis/Synthesis A10 10
MILLIPAK
The Millipak Final Filter is a membrane based filter that removes all particles and bacteria with a size greater than
0.22 m by a sieving mechanism. All contaminants larger than the filter pore size are retained on its surface. This
essentially will block and retain any bacteria present in the water flowing through it.
WHAT TYPES OF MILLIPAK FILTERS ARE THERE?
There are 2 types of Millipak Filters (sometimes called the Final Filter).
The Sterile Millipak 40
is generally used for some Life Science applications. This has a Millipore
Catalogue Number of MPGL04SK2. See Section 8-1 for more information.
The Non Sterile Millipak Express 20
is generally used for Analytical Science Applications including
applications where trace levels of Volatile Organic Compounds are desired or with Life Science
Applications where Endocrine Disrupter Analysis is being done. The Non Sterile Millipak Express 20
has a Millipore Catalogue Number of MPGP02001. See Section 8-1 for more information.
WHEN DO I CHANGE THE MILLIPAK?
The Millipak is changed when the Q-Gard Purification Pack and the Quantum Ultrapure Cartridge are changed.
However, it may be necessary to change the Millipak if it becomes clogged with particles or bacteria and reduces
the flowrate of Milli-Q Product Water.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Millipore milli-q and is the answer not in the manual?

Millipore milli-q Specifications

General IconGeneral
BrandMillipore
Modelmilli-q
CategoryWater Filtration Systems
LanguageEnglish

Related product manuals