MAINTENANCE
Milli-Q Synthesis/Synthesis A10 57
Chapter 6 MAINTENANCE
6-1 MAINTENANCE SCHEDULE
See the Maintenance Guide below for the typical maintenance that needs to be performed on your
Milli-Q Synthesis or Milli-Q Synthesis A10. The Catalogue Numbers and other ordering information can be found
in Sections 8-1 and 8-2. Detailed information about the various consumable items can be found in Sections 2-2
and 2-3.
Item or action Maintenance
needed
When? How to?
Q-Gard Pack
Replacement When the system resistivity or
TOC values are outside of
specification or the EXCH.
CARTRIDGE message is
displayed.
See Section 6-2
Quantum Cartridge
Replacement Replace when the Q-Gard Pack
is replaced.
See Section 6-3
Millipak
Replacement Replaced when the Quantum
Cartridge and Q-Gard Pack are
replaced or when the Product
Water Flowrate drops.
See Section 6-4
Sanitisation and cleaning of
the UF Cartridge
Cleaning Every 4 weeks. See Section 6-5
Screen Filter
Cleaning 2 times a year or as necessary. See Section 6-6
Flow Calibration
Perform a calibration
When the Millipak is
replaced.
After an AIR PURGE
After a ‘dirty’ UF Cartridge
is cleaned
Periodically.
See Section 4-11
POU Dispenser Arm
Tightening As necessary. See Section 6-7
A10 UV
Replacement Every 365 Days. See Section 6-8
A10 Cleaning
Perform a cleaning Automatic when the Quantum is
replaced.
----------------------------------
As necessary.
See Section 6-9
Air Purge of UF Cartridge
Perform an Air Purge Automatic when the Quantum is
replaced.
----------------------------------
As necessary.
See Section 6-10
UV Lamp
Replacement Every 365 Days See Section 6-11