USING THE MILLI-Q
Milli-Q Synthesis/Synthesis A10 56
Message and System
Status
What it means What to do
SYSTEM ERROR 8
The POU Dispenser will not dispense
water. You will not be able to
dispense water.
The electrical intensity being sent to the
UV Ballast Power Supply on the
Interface PC Board is < 5 mA or
> 950 mA. This is also the Power
Supply for the Solenoid Valve in the
POU Dispenser.
Contact Millipore.
SYSTEM ERROR 10
An error occurred with data transfer to
or from the Milli-Q microprocessor
(E
2
Prom).
Turn off the Milli-Q. Turn it back on
about 20 to 30 seconds later. If the
message continues, then contact
Millipore.
A10 ERROR 0
A10 PCB E
2
Prom defective. Turn off/on the Milli-Q and dispense
water for > 6 minutes. If the message
continues contact Millipore.
A10 ERROR 1
A10 PCB A/D converter defective. Turn off/on the Milli-Q and dispense
water for > 6 minutes. If the message
continues contact Millipore.
A10 ERROR 2
A10 Thermistor defective. Turn off/on the Milli-Q and dispense
water for > 6 minutes. If the message
continues contact Millipore.
A10 ERROR 3
Problem occurred with temperature
compensation.
Turn off/on the Milli-Q and dispense
water for > 6 minutes. If the message
continues contact Millipore.
A10 ERROR 4
The water entering the A10 is < 4 C. Contact Millipore.
A10 ERROR 5
The water entering the A10 is
> 41 C.
Contact Millipore.
A10 ERROR 6
The conductivity of the water entering
the A10 is > 1.1 S/cm.
Contact Millipore.
A10 ERROR 7
The temperature inside the A10 during
its analysis mode reached > 55 C.
Contact Millipore.
A10 ERROR 9
The TOC is very low or a problem
occurred with the A10.
Contact Millipore.