MAINTENANCE
Milli-Q Synthesis/Synthesis A10 72
6-7 HOW TO TIGHTEN THE POU DISPENSER ARM
The POU Dispenser Arm is moved up and down during the times that water is dispensed from the Milli-Q System.
Follow these instructions to tighten the POU Dispenser Arm
Locate the ‘Allen’ Wrench (6-sided black wrench) from the Accessories Bag (from the Shipping Box).
Locate the 2 bolts that hold the various pieces of the Dispenser Arm in place.
Tighten these bolts until the arm is secure.
You have finished tightening the POU Dispenser Arm.
6-8 HOW TO REPLACE THE A10 UV LAMP
The A10 UV Lamp is normally replaced once a year (every 365 Days).
A message, EXCHANGE A10 UV, will appear after 365 Calendar Days indicating that the A10 UV Lamp needs to
be replaced.
It is recommended to have a Millipore Field Service Representative change the A10 UV Lamp. The replacement of
the A10 UV Lamp involves removing the cover of the system. The instructions for replacing the A10 UV Lamp are
not included in this manual. The instructions are included with the replacement A10 UV Lamp.
Contact Millipore for a copy of these instructions if you would like to see them before ordering a replacement
A10 UV Lamp.
6-9 HOW TO CLEAN THE A10 TOC MONITORING DEVICE
The A10 TOC Monitoring Device may need to be cleaned sometimes. This cleaning does not involve any chemical
reagents. This cleaning is initiated with the Milli-Q software and is simple to perform.
HOW LONG DOES THE A10 CLEANING TAKE ?
The A10 Cleaning takes 60 minutes to complete. The LCD will show a countdown timer indicating the
remaining time until the A10 Cleaning is finished.
WHY WOULD I NEED TO DO AN A10 CLEANING ?
The A10 Cleaning is recommended when:
1. the Q-Gard and Quantum have been replaced (see the next question below).
2. the displayed TOC values are fluctuating.
3. the displayed TOC values are higher then normally seen.
4. the Milli-Q has been in STANDBY Mode or powered off for more than 12 hours.
WOULD AN A10 CLEANING EVER HAPPEN AUTOMATICALLY ?
An A10 Cleaning will occur automatically after the Q-Gard Pack and Quantum Cartridge are exchanged.
Normally, a 5 minute AIR PURGE Mode is started after the Quantum Cartridge is replaced. When the AIR PURGE
is finished, the 60 minute A10 CLEANING Mode is started.
See Section 4-7 for more information about the AIR PURGE and the automatic A10 CLEANING Mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Millipore milli-q and is the answer not in the manual?

Millipore milli-q Specifications

General IconGeneral
BrandMillipore
Modelmilli-q
CategoryWater Filtration Systems
LanguageEnglish

Related product manuals